NanoAssemblr® BLAZE™ and BLAZE+™

NanoAssemblr® BLAZE™ and BLAZE+™

To test the Safety and Efficacy in different species prior to clinical studies
Power to Scale 10mL - 10L

NanoAssemblr® BLAZE™ and BLAZE+™의 특징

  • Formulate 10mL-1L nanomedicines with BLAZE and up to 10L with BLAZE+
  • Complete control through dedicated computer system
  • Achieve flowrate up to 115ml/min to tune particle size
  • Inline dilution to eliminate contamination chances and ensure homegeneity
  • Maximum reproducibility
  • Optimal control for variation and human errors
  • Acquire scalability across NanoAssemblr platform

Features of NanoAssemblr® BLAZE™ and BLAZE+™

  • THE SCALABLE POWER OF NXGEN™

    THE SCALABLE
    POWER OF NXGEN™

    Advance candidates developed at bench scale by scaling up to the NxGen 400 and NxGen 500 cartridges—the same ones on the GMP System

  • EXPANDED PROCESS DEVELOPMENT CAPABILITIES

    EXPANDED PROCESS
    DEVELOPMENT CAPABILITIES

    Continuous flow pumping and in-line pressure monitoring provide an efficient platform for process development.

  • REPRODUCIBILITY

    REPRODUCIBILITY

    Exquisite batch-to-batch reproducibility enables the identification of critical process attributes from raw material supply to the final manufacturing steps.

  • INTUITIVE WORKFLOW

    INTUITIVE WORKFLOW

    Familiar interface, built-in checklists, saved recipes and complete run history ensures new users are quickly able to perform the studies they need.

TECHNICAL DESCRIPTION

Technical specifications
Electric power supply Voltage: 100 - 240 VAC,
Frequency: 50-60 Hz
Current: 1.2 A
Electrical input Voltage: 24 VDC
Current: 3.75 A
Environmental requirements Temperature: 5 – 35°C
Humidity: 0 – 40 % (non-condensing)
Pressure: 70 - 106 kPa
Ingress protection: IP20
Input/output connections Power cable: Power input
USB cable: Data input/output
Dimensions(WxDxH) With lid closed: 893 x 466 x 650 mm
With lid open: 893 x 511 x 1221 mm
Weight: 127*
Tolerances Flow rate accuracy: +/- 2.5% (average) of
the combined input lines (no-dilution)
Formulation pressure: Up to 150 psi
Volume accuracy: -4% to +2%

* The minimum weight the lab bench must support is 200kg

Parameter Allowable range
Volume 10 – 1000 mL
Reagent ratio 1:1 – 4:1
Total flow rate 4-115mL/min*
Dilution ratio 1:1 up to 4:1
(see working with recipes for more details)
Start waste 0 mL up to total volume
End waste 0 mL up to total volume